Optomet Nova Series laser Doppler vibrometer: non-contact velocity, displacement and acceleration, 1550 nm SWIR laser, DC-25 MHz. From Cromocol.

Get a Quote
Nova Series - Optomet

Product Description

The Optomet Nova Series is a single-point laser Doppler vibrometer from the CLASSIC Series. It measures the velocity, displacement and acceleration of a vibrating surface without contact, by evaluating the Doppler frequency shift of laser light reflected from the measurement point. Signal processing is fully digital and handled by the Optomet UltraDSP.

The measurement beam is a short-wavelength infrared (SWIR) laser at 1550 nm, paired with a visible 510-530 nm targeting laser for aiming. At this wavelength the system can emit up to 10 mW while remaining laser class 1, and the laser runs in a single longitudinal mode, so sensitivity does not depend on the measuring distance. Because the measurement is optical, vibrations can be recorded on surfaces that are difficult for visible-light vibrometers, such as deep-black, glass, carbon-composite and incandescent metal surfaces, without surface preparation.

The Nova Series covers a frequency bandwidth from DC to 25 MHz and vibrational velocities up to 30 m/s, with the available velocity, displacement and acceleration ranges set by the fitted decoder. Four interchangeable bayonet-mount lenses (Short-, Mid-, Long- and Super-Long-Range) provide working distances from 6 mm to 200 m. Digital output is over GBit-Ethernet and three BNC connectors provide analog output. The instrument is operated through a 3.5-inch color screen with a 20-segment LED bargraph, knobs, a touch screen and a key switch.

It is used for non-contact vibration measurement in materials research, automotive and aerospace component testing, civil engineering and quality control, including turbine-blade analysis and long-distance measurements where mounting sensors is impractical.

Specifications

  • Measured quantities: velocity, displacement, acceleration
  • Frequency bandwidth: DC to 25 MHz; vibrational velocity up to 30 m/s (decoder dependent)
  • Displacement resolution down to 50 femtometres
  • Measurement laser: 1550 nm SWIR (class 1, < 10 mW); target laser 510-530 nm (class 2, < 1 mW)
  • Interchangeable lenses with working distances from 6 mm to 200 m
  • Outputs: GBit-Ethernet (RJ45) and 3 x BNC analog
  • Dimensions 380 x 180 x 148 mm (excluding handle and lens); weight 8 kg
  • Power: 110-240 V AC (50-60 Hz) or 12 V DC; 27 W

Technical Specifications


Measurement
  • Measured quantities: velocity, displacement, acceleration
  • Max. frequency bandwidth: DC to 25 MHz
  • Max. vibrational velocity: up to 30 m/s (decoder dependent)
  • Velocity measuring ranges: up to 15
  • Displacement measuring ranges: up to 19
  • Acceleration measuring ranges: up to 15
  • Displacement resolution: down to 50 fm
  • Signal processing: digital (Optomet UltraDSP)

Optics & Laser
  • Measurement laser: SWIR, 1550 nm (class 1, < 10 mW)
  • Target laser: 510-530 nm (class 2, < 1 mW)
  • Single longitudinal laser mode: sensitivity independent of measuring distance
  • Lens mount: bayonet-lock, interchangeable
  • Short-Range (SR) lens: working distance 6 mm to 5 m; spot size 25 µm @ 6 mm
  • Mid-Range (MR) lens: working distance 125 mm to 10 m; spot size 43 µm @ 125 mm
  • Long-Range (LR) lens: working distance 450 mm to 100 m; spot size 67 µm @ 450 mm
  • Super-Long-Range (SLR) lens: working distance 2,300 mm to 200 m; spot size 245 µm @ 2,300 mm
  • Focus: autofocus, remote or manual

Decoders
  • D-VD-0N (Start): 7 velocity ranges, 2.5 m/s max., DC to 100 kHz
  • D-VD-5N-30 (Master++): 15 velocity ranges, 30 m/s max., DC to 25 MHz
  • Displacement decoder D-DD-5N-30: 19 ranges (optional)
  • Acceleration decoder D-AD-5N-30: 15 ranges (optional)
  • Decoder upgrades available remotely without factory return

Interfaces
  • Digital output: GBit-Ethernet (RJ45)
  • Analog output: 3 × BNC
  • Display: 3.5" color screen with 20-segment LED bargraph
  • Input: touch screen, knobs with push-button, key switch

System
  • Dimensions: 380 × 180 × 148 mm (excluding handle and lens)
  • Weight: 8 kg
  • Mounting: standard tripod screw thread
  • Power supply: 110-240 V AC (50-60 Hz) or 12 V DC
  • Power consumption: 27 W
  • Operating temperature: 0 °C to 40 °C

Frequently Asked Questions

What is the Nova Series and what does it measure?
The Nova Series is a single-point free-beam laser Doppler vibrometer in Optomet's CLASSIC Series. It measures velocity, displacement and acceleration of a vibrating surface without contact, using the Doppler shift of laser light reflected from the measurement point. It covers a frequency bandwidth from DC to 25 MHz and vibrational velocities up to 30 m/s, depending on the decoder fitted.
How does the 1550 nm laser affect measurement?
The Nova Series uses a short-wavelength infrared (SWIR) measurement laser at 1550 nm together with a visible 510-530 nm targeting laser. At 1550 nm the system can emit up to 10 mW while remaining laser class 1, and photodetectors in this range operate at high efficiency. This raises the available signal level compared with visible HeNe-based systems. The laser operates in a single longitudinal mode, so sensitivity does not change with measuring distance.
What surfaces and distances can it measure?
Because measurement is optical, the Nova Series records vibrations on surfaces that are difficult for visible-light vibrometers, including deep-black, glass, carbon-composite and incandescent metal surfaces, without surface preparation. With the appropriate lens, working distances extend from 6 mm up to 200 m.
What lens options are available?
Four interchangeable bayonet-mount lenses are offered: Short-Range (6 mm to 5 m), Mid-Range (125 mm to 10 m), Long-Range (450 mm to 100 m) and Super-Long-Range (2,300 mm to 200 m). Each supports autofocus, remote and manual focusing, with laser spot sizes from 25 micrometres at the shortest working distance.
How are the outputs and decoders configured?
The instrument provides a GBit-Ethernet (RJ45) digital output and three BNC analog outputs, with signal processing handled by the digital Optomet UltraDSP. Velocity, displacement and acceleration decoders define the available measuring ranges; for example the D-VD-0N covers 7 velocity ranges to 2.5 m/s, while the D-VD-5N-30 covers 15 ranges to 30 m/s. Decoders can be upgraded remotely without returning the unit to the factory.
What are typical applications?
The Nova Series is used for non-contact vibration measurement in materials research, automotive and aerospace component testing, civil engineering and quality control. Common tasks include turbine-blade analysis, measurements on lightweight or non-cooperative materials, and long-distance measurements where attaching sensors is impractical.

Request a quote for Nova Series - Optomet

Fill out the form, give us a call or send an email and we will contact you as soon as possible.

Back to all products